Treatment of Iatrogenic Renal Artery Perforation with a Covered Stent and Subsequent Rheolytic Thrombectomy

  • Fri, 9/5/08 - 3:36pm
  • 0 Comments
  • 2885 reads
Author(s): 

Jeffrey M. Friedel, MD, Ramzi Khalil, MD, David Lasorda, MD

Introduction
Iatrogenic renal artery perforation is a rare, but well-described complication of percutaneous renal artery intervention.1-3 Treatment has traditionally involved surgical intervention including renal artery reconstruction, bypass grafting, and even nephrectomy.4,5 More recently, several non-surgical techniques have been described, including balloon tamponade, custom-made stent graft deployment, and use of an autologous vein-covered stent.6,7 The JoStent® (JoMed, Rancho Cordova, California) has been approved by the FDA under humanitarian device exemption for the treatment of coronary artery perforations. Although experience in a variety of clinical settings has been described,8 use of the Jostent for renal artery perforations has not been reported. We describe a case in which we successfully treated a renal artery perforation with balloon tamponade and deployment of the Jostent, with subsequent rheolytic thrombectomy using the AngioJet thrombectomy catheter (Possis Medical, Minneapolis, Minnesota).

Case Report
The patient is a 73-year-old female with a past medical history significant for coronary artery disease, hypertension, diabetes mellitus type II, chronic renal insufficiency, and morbid obesity. In addition, she had known right renal artery stenosis (documented by angiography) that was managed medically. The left renal artery was very small and supplied an atrophic left kidney. Over the past year, her hypertension became increasingly resilient to medications and the decision was made to perform percutaneous right renal artery intervention.

An 8 Fr sheath was inserted in the right femoral artery and and the right renal artery was engaged with an 8 Fr RDC catheter (Cordis Corporation, Miami Lakes, FL). Selective angiography revealed an 80% lesion involving the ostium of the right renal artery (Figure 1). The patient received 5,000 units of heparin, and a 190 cm .014 inch S'port™ wire (Guidant Corporation, Santa Clara, CA) was advanced across the lesion into the distal renal artery. The lesion was directly stented with a 6.0 mm x 24 mm Genesis™ stent (Cordis Corp.) to 12 atmospheres, and subsequently post-dilated with a 7.0 mm x 15 mm Viatrac® balloon (Guidant Corp.) to 8 atmospheres. Final angiography revealed a 0% residual lesion (Figure 2). The patient's activated clotting time at the end of the procedure was 285 seconds.

Approximately 2 hours after the intervention, the patient developed severe right flank and back pain. She was emergently returned to the angiography suite. The right renal artery was again engaged with an 8 Fr RDC catheter. The artery was widely patent with normal flow but there was an area of dye extravasation distal to the previously placed stent (Figure 3). Protamine was not administered given the concern of anaphylaxis in this patient who previously received protamine-containing insulin.

Consideration was given to placing a large (> 5 mm) perfusion balloon or covered stent; however, attempts at engaging the right renal artery with multiple larger guides were unsuccessful. We re-engaged the artery with the 8 Fr RDC catheter. A 190 cm .014 inch Marker wire (Cordis Corp.) was advanced into the distal renal artery. A 4.5 mm x 19 mm Jostent (the largest such covered stent in our supply) was advanced across the area of extravasation and inflated to 16 atmospheres. The Jostent was post-dilated with a 6.0 mm x 15 mm Viatrac balloon to 12 atmospheres. The post-deployment angiogram showed significant improvement in the perforation but a small leak persisted. A 4.0 mm x 20 mm ACS Esprit™ perfusion balloon (Guidant Corp.; the largest in our supply) was then advanced across the stented area and inflated to 8 atmospheres. Given that the right kidney was most likey the sole functioning kidney, the patient was sent for a CT scan of the abdomen and pelvis to rule out any significant organ compression that may require surgical intervention. The scan revealed the presence of a moderate-sized retroperitoneal bleed which did not compress the renal artery or kidney.

The patient was returned to the angiography suite, and the perfusion balloon was deflated. Repeat angiography confirmed that the perforation had been sealed. Unfortunately, there was now a large, mobile filling defect within the artery consistent with thrombus (Figure 4). A 300 cm 0.014 inch S'port wire was advanced into the distal artery. A 4 Fr AngioJet catheter was passed through the area of thrombus and there was complete resolution of the filling defect with excellent antegrade flow in the renal artery (Figure 5). A total of 325 ml of contrast was utilized, and over the next few days, her creatinine increased from a baseline of 1.9 to 3.2. This subsequently improved to 2.1 prior to discharge. Her hypertension was well-controlled with beta blockers, hydralazine, and clonidine. The patient was discharged home in good condition 6 days after the initial procedure.

Discussion
Percutaneous intervention has now become a common practice in the treatment of renal artery stenosis, with efficacy and safety similar to or better than surgery.9 Advances in both interventional equipment and operator technique have lowered complication rates over the past decade.3 Despite these advances, the risk of renal artery perforation still exists. Perforation can occur from the guidewire (especially hydrophilic wires), angioplasty balloon, stent, and even from the guiding catheter itself. In our case, the perforation was most likely secondary to direct placement of the stent, as this was the site of dye extravasation and there was minimal difficulty with initial wiring of the vessel. Quantitative measurements of the vessel, or pre-dilatation prior to stent placement, may have helped to avoid this complication.

References: 

1. Henry M, Amor M, Henry I, et al. Stents in the treatment of renal artery stenosis: Long-term follow-up. J Endovasc Surg 1999;6:42–51.
2. Mahler F, Triller J, Weidmann P, Nachbur B. Complications in percutaneous transluminal dilatation of renal arteries. Nephron 1986;44(Suppl 1):60–63.
3. Zeller T, Frank U, Muller C, et al. Technological advances in the design of catheters and devices used in renal artery interventions: Impact on complications. J Endovasc Ther 2003;10:1006–1014.
4. Dean RH, Callis JT, Smith BM, Meacham PW. Failed percutaneous transluminal renal angioplasty: Experience with lesions requiring operative intervention. J Vasc Surg 1987;6:301–307.
5. McCann RL, Bollinger RR, Newman GE. Surgical renal artery reconstruction after percutaneous transluminal angioplasty. J Vasc Surg 1988;8:389–394.
6. Morris CS, Bonnevie GJ, Najarian KE. Nonsurgical treatment of acute iatrogenic renal artery injuries occurring after renal artery angioplasty and stenting. Am J Roentgenol 2001;177:1353–1357.
7. Bates MC, Shamsham FM, Faulknier B, Crotty B. Successful treatment of iatrogenic renal artery perforation with an autologous vein-covered stent. Catheter Cardiovasc Interv 2002;57:39–43.
8. Gercken U, Lansky AJ, Buellesfeld L, et al. Results of the Jostent coronary stent graft implantation in various clinical settings: Procedural and follow-up results. Catheter Cardiovasc Interv 2002;56:353–360.
9. Geroulakos G, Missouris C, Mitchell A, Greenhalgh RM. Endovascular treatment of renal artery stenosis. J Endovasc Ther 2001;8:177–185.
10. Schachinger V, Hamm CW, Munzel T, et al. A randomized trial of polytetrafluoroethylene membrane-covered stents compared with conventional stents in aortocoronary saphenous vein grafts. J Am Coll Cardiol 2003;42:1360–1369.
11. Allaire E, Melliere D, Poussier B, et al. Iliac artery rupture during balloon dilatation: What treatment? Ann Vasc Surg 2003;17:306-314. Epub 2003 Apr 28.
12. Gunning MG, Williams IL, Jewitt DE, et al. Coronary artery perforation during percutaneous intervention: Incidence and outcome. Heart 2002;88:495–498.
13. Bates MC, Shamsham FM, Faulknier B, Crotty B. Artery perforation during catheterization: Fighting with a catastrophe. Catheter Cardiovasc Interv 2002;57:44–46.
14. Baltacioglu F, Cimsit NC, Cil B, et al. Endovascular stent-graft applications in latrogenic vascular injuries. Cardiovasc Intervent Radiol 2003;26:434–439.
15. Fukutomi T, Suzuki T, Popma JJ, et al. Early and late clinical outcomes following coronary perforation in patients undergoing percutaneous coronary intervention. Circulation 2002;66:349–356.
16. Kuntz RE, Baim DS, Cohen DJ, et al. A trial comparing rheolytic thrombectomy with intracoronary urokinase for coronary and vein graft thrombus (the Vein Graft AngioJet Study [VeGAS 2]). Am J Cardiol 2002;89:326–330.
17. Rinfret S, Katsiyiannis PT, Ho KK, et al. Effectiveness of rheolytic coronary thrombectomy with the AngioJet catheter. Am J Cardiol 2002;90:470–476.
18. Singh M, Tiede DJ, Mathew V, et al. Rheolytic thrombectomy with Angiojet in thrombus-containing lesions. Catheter Cardiovasc Interv 2002;56:1–7.
19. Ansel GM, George BS, Botti CF, et al. Rheolytic thrombectomy in the management of limb ischemia: 30-day results from a multicenter registry. J Endovasc Ther 2002;9:395–402.
20. Greenberg JM, Steiner MA, Marshall JJ. Acute renal artery thrombosis treated by percutaneous rheolytic thrombectomy. Catheter Cardiovasc Interv 2002;56:66–68.

  • 1
  • 2
  • next ›
  • last »
image description image description

Post new comment

  • Web page addresses and e-mail addresses turn into links automatically.
  • Lines and paragraphs break automatically.

More information about formatting options

Image CAPTCHA
Enter the characters shown in the image.

  • Advertise your Job Here
    For information on posting classified ads, please contact:
    Alex Dulnikowski, Classified Sales Manager
    (800) 237-7285, ext. 205

vdm Blogs

PROTECT carotid stent trial provides further evidence that CAS is getting better and becoming more competitive with CEA

Frank J Criado MD FACS FSVM

A New Algorithm to Treat Patients with Peripheral Vascular Disease

Robert S. Dieter MD RVT and Aravinda Nanjundappa MD RVT

In-Stent Restenosis in the SFA Remains a Significant Unresolved Problem

Frank J Criado MD FACS FSVM

Support Comes From Many Directions

Richard R. Heuser MD FACC FACP FESC FSCAI

Pedal Artery Access: Advances in Management of Critical limb ischemia

Robert S. Dieter MD RVT and Aravinda Nanjundappa MD RVT
more »

Vascular Newswire

  • Medtronic Stent Resulted in 90% Freedom from Reinterventions in Narrowed Leg Arteries at 12 Months in International Study
    Thu, 02/09/12 - 10:52am
  • AngioDynamics Launches DuraFlow 2 Chronic Hemodialysis Catheter
    Wed, 02/08/12 - 10:18am
  • Robotic-Assisted Pci Procedures Using Corpath 200 System Will Be Discussed During CRT2012 Conference
    Tue, 02/07/12 - 10:00am
  • Vascular Solutions Launches Reprocessing Service For Closurefast Vein Catheters
    Fri, 02/03/12 - 1:14pm
more »

Clinical Events Calendar

  • American Venous Forum 24th Annual Meeting
    Wed, 02/08/2012 - Sat, 02/11/2012
    Orlando, FL, United States
  • JIM 2012
    Thu, 02/09/2012 - Sat, 02/11/2012
    Rome, Italy
  • Cardiovascular Care Update 2012 (CVC)
    Fri, 02/10/2012 - Sat, 02/11/2012
    Scottsdale, AZ, United States
more »

Poll

How do you feel about the continued practice of screening patients for asymptomatic CAD?:
REVIEW OUR OTHER Cardiology BRANDS

Our other resources for healthcare professionals.

HMP Communications © 2012 HMP Communications

HMP Communications LLC (HMP) is the authoritative source for comprehensive information and education servicing healthcare professionals. HMP’s products include peer-reviewed and non-peer-reviewed medical journals, national tradeshows and conferences, online programs and customized clinical programs. HMP is a wholly owned subsidiary of HMP Communications Holdings LLC. ©2012 HMP Communications