The Role of Embolic Protection Devices in Renal Angioplasty and Stenting
Clinical Review:
The Role of Embolic Protection Devices in Renal Angioplasty and Stenting

- 1,2Michel Henry, MD, 3Isabelle Henry, MD, 4Christos Klonaris, MD, 5Antonio Polydorou, MD, 6Amanda Polydorou, MD, 1,2Michele Hugel, MD

Purpose. A renal artery stenosis is frequent and usually caused by atherosclerosis. Renal artery angioplasty and stenting (RAAS) is now the first treatment to be proposed, giving good immediate and long-term results. But one concern is the postprocedural deterioration of the renal function, which may occur in 20–40% of the patients, and therefore limits the immediate benefits of the technique. Atheroembolism seems to play an important role. We postulated that the use of renal embolic protection devices could reduce the risk of renal embolism and avoid deterioration of the renal function. Methods. In 105 hypertensive patients (M:72), mean age: 65.4 ± 11.7 years (22–87) with atherosclerotic renal artery stenosis (18 bilateral), 124 RAAS were performed under protection. Nine patients had solitary kidneys and 38 had renal insufficiency. We used occlusion balloons (n = 46) or filters (n = 78), which allow a continuous flow. We recently experimented with a new filter (FiberNet®), which can capture particles of 30–40 microns without compromising the flow. Generated debris were removed and analyzed. Blood pressure and serum creatinine levels were followed. Results. Immediate technical success was 100%. Of the lesions, 96/124 were stented directly. Visible debris were aspirated with Percusurge from all patients and removed with filters in 80% of the cases (100% with FiberNet). Mean particle number: 98.1 ± 60.0. Mean diameter: 201 ± 76 (38–6206). Mean occlusion time: 6.55 ± 2.46 min (Percusurge). Mean time in situ (filters): 4.2 ± 1.1 min. With the FiberNet, five times more particles were removed. We observed one acute renal function (RF) deterioration. Mean follow-up: 18.2 ± 8 months. Mean creatinine level remained constant during follow-up. At 6 months follow-up in 91 patients, 69 patients had stabilized, 21 with baseline renal insufficiency improved and there was only one RF deterioration (1.1%) in a patient with moderate renal insufficiency. At 2 years, 54 patients were stabilized, 19 improved and we had only 2 RF deterioration (3%). Conclusion. The preliminary results suggest the feasibility and safety of distal protection during renal interventions to protect against atheroembolism and consequent deterioration of renal function after the procedure. The beneficial effects of this technique should be evaluated in randomized studies.


Introduction
Atherosclerotic renovascular disease is increasingly recognized thanks to technical improvements in duplex ultrasound, magnetic resonance angiography, CT scan, routine renal angiography during cardiac catheterization, coronary procedures and particularly in hypertensive or multivascular diseased patients. It represents an important public health problem.

A renal artery stenosis (RAS) is usually caused by atherosclerosis (80% of cases in patients over 40 years), and in rare cases is due to fibromuscular dysplasia (10% of cases and more often in young patients), arteritis (Takayasu’s disease), neurofibromatosis and radiation. It can also be diagnosed in a renal transplant or in a renal bypass graft.

The prevalence of RAS is high. Rihal found RAS greater than 50% in 19.2% of patients during cardiac catheterization of 297 hypertensive patients.1 The prevalence of RAS is 35–45% in patients with peripheral vascular disease, 14–24% in patients with cerebrovascular disease and 7–30% in patients with coronary heart disease.2–4 In patients with renal insufficiency, the incidence of unsuspected RAS is as high as 24%.5 RAS greater than 60% has been reported to be 6.8% in patients older than 65 years of age.6

The natural history of RAS depends on its etiology. Atherosclerotic RAS, as well as renal artery stenosis due to Takayasu’s Arteritis, have a high tendency to progress with time, resulting in renal artery occlusion, loss of renal mass and a subsequent decrease in kidney function. In the vessels with the most severe atherosclerotic stenoses, the result is total occlusion in 16% of the patients.7–11 On the other hand, RAS due to fibromuscular dysplasia rarely tends to progress to occlusion.12,13

Atherosclerotic RAS can lead to different clinical situation:
• A renovascular hypertension (secondary hypertension), which accounts for 1–5% of all cases of hypertension.14,15
• A renal insufficiency. A rise in serum creatinine following initiation of antihypertensive therapy with ACE inhibitors may lead to the diagnosis of RAS. A RAS may be severe enough to cause ischemia and tissue damage, as is often shown by asymmetry in renal size.13
• A flash pulmonary edema is often the first clinical symptom of bilateral RAS.16

Renovascular hypertension can also be caused by non-atherosclerotic RAS such as fibromuscular dysplasia, which occurs in younger patients (< 40 years of age). It is probably the most common cause of hypertension in young children17,18 and has a higher prevalence in females.

Atherosclerotic renovascular disease represents an important public health problem. It has been demonstrated that it increases cardiovascular and all-cause mortality.

Aggressive treatment of a severe RAS (defined as a diameter stenosis of at least 70%) is recommended for patients with uncontrolled hypertension, renal insufficiency, congestive heart failure, unstable angina and for patients with a solitary or a single-functioning kidney. The treatment of RAS without hypertension or renal insufficiency is debatable but could be considered with an intention of preserving renal function and renal artery patency.

The treatment options for a RAS include medical therapy, balloon angioplasty (with and without stenting), and surgery. Surgery carries a significant risk with a 2–7% perioperative mortality rate and a 17–31% morbidity rate. In addition, deterioration in renal function occurs in 11–31% of patients, reocclusion and restenosis in 5–18%.24–26 Indications for surgery should be limited to failed percutaneous approach, hostile aorta, or infrarenal total occlusion (in association with aortic surgery).

Percutaneous transluminal renal angioplasty (PTRA) has become the cornerstone of therapy for addressing RAS and is now the first treatment to be proposed. Balloon angioplasty alone was initially proposed and is still the first line of therapy for fibrodysplasic RAS. However, several authors have reported the successful use of endovascular stents for treating suboptimal angioplasty results, and as a primary intervention for atherosclerotic lesions (particularly ostial lesions) with better immediate and long-term results than with PTA alone.27–32

Renal angioplasty stenting can be performed using the femoral approach in most cases. The brachial or radial approach can also be used. The technique benefits from the improvements of coronary technique are monorail systems for balloon and stents, low profile devices, 0.014'' or 0.018'' guidewires, allowing direct stenting in 80–90% of the procedures.

Procedural success is excellent (98–100%), with a low complication rate, a low restenosis rate and a good long-term patency rate of 85–98%.21–26

The benefit of PTRA includes complete cure (7–19%), or at least easier management of hypertension (52–74%) in addition to preservation, or improvement of renal function.27,28,31 However, post-procedural deterioration of the renal function occurs in a subset of patients after PTRA.32–34 We hypothesize that atheroembolism during the procedure is a precipitating factor for this complication and a major preventable factor limiting the benefits derived from PTRA. This hypothesis is supported by recent studies.35 In order to eliminate or reduce the risk of atheroembolic material being carried into the renal parenchyma, we applied the technique of distal embolic protection, using balloon or filters positioned distal to the lesion, a technique currently developed and approved for use in the coronary and cerebral circulations.36–38 Use of such devices during PTRA may increase the rate of beneficial clinical responses after revascularization.

We will report a series of 124 RAS done under protection in 105 patients. All patients provided written informed consent.


Effects on Renal Function: Rationale for Protection Devices
The effects of renal angioplasty and stenting on renal function are controversial. There is no randomized study comparing either balloon angioplasty or stenting with medical therapy. The meta-analysis reported by Nordmann et al39 demonstrated no consistent difference in changes in the renal function.
According to recent American Heart Association Guidelines,40 a slowed decline in renal function is sufficient to support the claim that renal artery angioplasty is beneficial. Recent studies regarding the effects of peripheral renal angioplasty or stenting on renal function showed that a large percentage of patients seem to benefit from the procedure with a stabilization or improvement in renal function (Table 1). Renal stenting in selected patients could slow the progression of renovascular renal failure.41–45,50–58

In patients with a normal baseline creatinine level, renal function was mostly preserved.31,43,50,54
In a non-randomized study, PTRA improved renal function in 41–43% of patients.59

Blum et al found no significant change in creatinine independent of baseline renal function.30

White et al,54 in his series of 100 patients, found no significant change in the creatinine level. However, in 9 of 44 patients with renal insufficiency, creatinine normalized.

The effect of PTRA on renal function was demonstrated by La Batatide-Alanore et al.60 After following 32 patients who underwent PTRA for a mean of 6 months, split renal function was completed using Captopril renal scintigraphy. The renal function of the treated kidney was found to improve significantly after PTRA. This suggests that PTRA improves the renal function in patients with unilateral RAS.

There are few data on patients with severe renal dysfunction. In Zeller’s series,50 renal function improved in 71% of patients with severe renal insufficiency, 21% were unchanged, and 8% deteriorated. In 32%, planed chronic hemodialysis was deferred. In contrast, Nama et al61 showed no improvement 1 year after the procedure in 40 azotemic patients with 61 RAS. In 60% of these patients, renal function was improved or preserved, and in 40% it was deteriorated. There was no difference found between bilateral and unilateral disease.


Despite these generally favorable results in many published series, a decline in renal function was noted in a subset of patients even after successful initial technical results and a good long-term patency of the renal artery. Deterioration in renal function may be seen in 20–30% of the patients (Table 1) or even more. In most of the series this depends on the renal function and the creatinine level at baseline.32,34,42,44 Dorros et al32 reported a deterioration of the renal function in 47% of the patients with a creatinine level of over 2 mg/dl.

Subramanian et al62 demonstrated a worsening in renal function in 24% of non-diabetic patients and 27% of diabetic patients with renal insufficiency and Guerrero et al47 reported similar results in 31% of patients with renal insufficiency.

Zeller et al50 recently reported the long-term impact of stent-supported angioplasty on renal function in a series of 340 hypertensive patients. During a mean follow-up period of 34 ± 20 months, serum creatinine significantly decreased from 1.45 ± 0.87 to 1.39 ± 0.73 mg/dl (p = 0.048). The renal function was improved in 34% of the patients, remained unchanged in 39% but deteriorated in 27%. Baseline serum creatinine, bilateral intervention, percent diameter stenosis and triple vessels coronary diseases were independent predictors of improved renal function during follow-up. The serum creatinine decrease was not significant in patients with diabetes mellitus.

Other reports41,43,55 suggest that only treatment of bilateral, severe RAS is beneficial to renal function.

In contrast to other series, the highest proportion (36%) of Zeller’s patients with deteriorated renal function was found in the subgroup with normal baseline serum creatinine. However, in 90% of these patients, the increase in serum creatinine concentration was within the normal range. The more severe the renal dysfunction was at baseline, the more the patients benefited from the intervention. Improved, or unchanged, mean serum creatinine concentrations were seen in 64% of patients with normal renal function, 82% of patients with moderate renal impairment and 92% of those with severely impaired renal function. Improvement in renal function was also seen in 37% of the patients with nephrosclerosis and a resistance index of more than 0.8.

Watson et al58 assessed the effect of renal artery stenting on renal function in patients with chronic renal insufficiency (creatinine > 1.5 mg/dl) by comparing the slopes of the regression lines derived from the reciprocal of serum creatinine versus time plotted before and after stent deployment. Before stent deployment, all patients exhibited a negative slope indicating progressive renal insufficiency. After stent deployment the slopes were positive in 18 patients (72%), indicating improvement in renal function and less negative in 7 patients (28%) indicating a persistence of decline in renal function deterioration, albeit at a lower rate than before the procedure in these 7 patients.

Many factors may be responsible for this functional deterioration: contrast media induced nephrotoxicity, progression of concomitant nephrosclerosis, lesion recurrence, hyperperfusion syndrome, and glomerular injury. However, atheromatous embolism seems to play an important role and is an increasingly recognized cause of renal function deterioration. It is demonstrated that atherosclerotic debris commonly embolize from lesions in many vascular territories during percutaneous intervention.63 Evidence of distal embolization as a frequent complication was first seen in saphenous vein graft intervention.64 Atheroembolism is also proven during catheter treatment of certain native coronary lesions,65 and during carotid and renal stenting.36–38,66,67 Distal protection devices have been able to recover embolic debris in all these territories and significantly reduce its incidence of these complications.

During a renal angioplasty, cholesterol atheromatous embolism is caused by the release of microscopic plaque fragments and cholesterol crystals from the renal artery lesion or the atherosclerotic aorta into parenchymal renal vasculature during the procedure. Instrument manipulation in the aorta and renal arteries can result in detachment and embolism of atheromatous debris from ulcerated plaques. The large size of the devices used, increased length or specific difficulties of the procedure may be contributory. Walker et al67 recently demonstrated the great potential for embolic debris during the placement of the guiding catheter, sheath or diagnostic catheter. He performed an aggressive aspiration of the guiding catheter or sheath before any contrast injection. Large particles of atherosclerotic debris (1–3 mm) were discovered in 41.7% of the patients. He proposed that careful aspiration of catheters before injections or interventions should routinely be performed. Patients with severe atheromatous disease of the aorta and its branches, ulcerated plaques, associated lesions, such as aneurysm, and dissection are candidates for these complications.

Hiramoto et al35 demonstrated that angioplasty and stenting of ex-vivo aortorenal atheroma specimens using 0.018-inch guidewire system was associated with thousand of atheroemboli. Each manipulation of the specimens including simply advancing the guidewire through the lesion released thousand of fragments. The number of fragments in each size category increased with decreasing particle size. Positioning and deploying the stent released an additional bolus of fragments similar to that released after balloon angioplasty. All these fragments were of sufficient size to create vascular occlusions and initiate significant renal parenchyma damage. This author concludes that the results of angioplasty procedures could be improved by placing distal protection devices to prevent atheroembolization.

Even with modern techniques, use of a low profile system cannot prevent atheroembolism and renal function deterioration as was recently reported by Nolan et al,68 who published a series describing 82 procedures with a renal function deterioration in 24% of the patients at 1 year.

The true incidence of atheroembolism is uncertain. Many patients can have a silent course because of the large functional kidney reserve, which allows normal serum creatinine values despite a significant decline in total glomerular filtration capacity. Therefore, only the most severe cases may be detected, especially in patients with preprocedural renal dysfunction and limited functional reserve.

Abnormal serum creatinine may only be observed if 50% of the nephron population is destroyed. Most patients reach a peak serum creatinine level over 3 to 8 weeks, but onset can also be sooner.

Few studies have addressed the problem of atheroembolism following RAS. Boisclaire et al reported 4 cases in 33 procedures.34 Van de Ven et al27 reported 2 cases in 24 procedures inducing renal insufficiency. Commeau et al69 reported acute deterioration of renal function due to cholesterol embolism necessitating hemodialysis. Morice et al70 also published 2 cases of partial renal infarctions in a series of 80 patients despite the use of a new low profile device and direct stenting. In the Aspire 2 Study,71 major embolic events occurred in 6.3% of the procedures.

The clinical manifestation of atheroembolism may be non-specific.
• Thadhani et al72 reported a series of 52 patients with both renal failure and histologically proven atheroembolism after angiography in cardiovascular surgery. Within 30 days of their procedure, 50% of the patients had cutaneous signs, 14% blood eosinophilia.
• Most patients reach a peak serum creatinine level over 3 to 8 weeks73 but onset may be sooner.31
• Haggie et al74 reported 4 patients with proteinuria and nephritic syndrome.

Atheroembolism can lead to different degrees of renal impairment:
- Moderate less of renal function;
- Severe renal failure requiring dialysis;
- Abrupt and sudden onset of renal failure;
- More frequently a progressive loss of renal function over 3 to 8 weeks;
- Chronic stable and asymptomatic renal insufficiency.

The diagnosis of atheroembolism is difficult. There is no specific test. Renal biopsy is the only definitive diagnostic tool, but its routine application is problematic. For these reasons, atheroembolism after renal intervention is often misdiagnosed as dye-induced nephrotoxicity or the progression of nephrosclerosis. However nephrotoxicity due to contrast media generally appears within 1 or 2 days or weeks. A prevention is essential in patients with renal insufficiency or in high risk patients (diabetes, elderly patients), the use of gadolinium, CoÇ angiography may be useful to reduce this risk.

Renal atheroembolism not only poses a risk of renal function deterioration but also seems to decrease survival in patients undergoing endovascular procedures for RAS. Krishnamurthi et al74 evaluated its impact on survival in 44 patients who had surgery for atherosclerotic RAS and concomitant intraoperative renal biopsy for detection of atheroemboli. Atheroembolic disease was identified in the biopsy specimens in 16 (36%) patients and correlated significantly with decreased survival (54% 5-year survival in this group versus 85% in patients without atheroembolism, p = 0.011).

Boero et al75 recently pointed out the bad prognosis of renal atheroembolism. In a series of 22 patients with an onset of symptoms from a few hours to 60 days, 11 patients (50%) were put on dialysis with a partial functional recovery in 4, while 11 patients (50%) died.

Thus, it can be concluded that cholesterol embolism is a frequent cause of renal failure and is associated with a high mortality rate. The increasing number of such patients, the cost due to renal function deterioration and subsequent end stage renal disease requiring dialysis represents a significant long-term problem.
No specific treatment can be suggested for renal atheroembolism. Therefore, the main aim should be the prevention during renal interventions. The selection of the patients may limit the risk, but more and more elderly, high-risk patients with advanced atheromatous diseases need treatment, and it is difficult to refuse these patients the benefits of the procedure.

Certain technical points are very important and need to be mentioned. The procedures should be as atraumatic as possible with use of small devices and the adaptation of coronary angioplasty techniques. Direct stenting is not sufficient to avoid embolism. As mentioned, Walker et al proposed the careful aspiration of the catheters. The “no touch” technique was also proposed to minimize atheroembolization.76
Beyond these technical considerations to circumvent atheroembolisms, we applied the concept of protected renal angioplasty and stenting.36–38

The rationale for distal embolic protection is similar to that of brain protection during angioplasty and stenting of the carotid arteries. Several studies have shown that protection devices with occlusion balloon or filter are effective in reducing the risk of embolization to the brain77,78 and that these techniques are now mandatory in this field and represent the standard of care.79 We postulated that the same technique could be suitable during renal angioplasty and stenting to reduce the risk of atheroembolism and deterioration of the RF.

Impact of Protection Devices during Angioplasty of RAS
Techniques. The technique has been previously described.36–38 All procedures are performed under local anesthesia and intravenous sedation. A 6–8 Fr guiding catheter is placed at the ostium of the renal artery, most of the time via a percutaneous femoral approach.
Two types of protection devices can be used:
• Occlusion balloon. The Guardwire system (Medtronic, Minneapolis, Minnesota) is the same as that used for carotid protection.
• Filters. Current filters available on the market can be placed in the renal artery (same filters as for carotid angioplasty) (Figure 1).

With the guiding catheter appropriately positioned, the protection device is carefully advanced across the lesion and positioned 2 or 3 cm beyond the target site. In the majority of the cases, crossing the lesion with a protection device is easy and is the same technique as with any guidewire.

In a few cases (angulated artery, tight, calcified stenosis), it could be helpful to first place a more supportive “buddy wire” in the renal artery before attempting to cross the lesion with the protection device. In case of very tight stenosis, a predilatation of the lesion with a coronary balloon may also be indispensable before crossing the lesion.

With a Percusurge device, the Microseal adapter is attached and the occlusion balloon inflated to occlude the renal artery. On detaching the adapter, the occlusion balloon remains inflated. With a filter, it is deployed by carefully removing the delivery sheath. Contrast injection confirms the accurate positioning of the protection device. Angioplasty and stenting of the renal artery can be performed under protection using low profile devices, monorail technique. We perform a predilatation of the stenosis only in case of tight or calcified lesions. A direct stenting is done in the majority of the cases.

Different types of stents were implanted. The diameter of the stent is chosen equal to the diameter of the artery. The stent is carefully positioned so that it protrudes into the aortic lumen by 1 or 2 mm.
After stent placement, the protection device is retrieved:

• If a Percusurge device is used, after stent deployment the aspiration catheter is advanced over the wire to the level of the lesion and positioned adjacent to the protection balloon. Any debris is removed using a 20 ml syringe connected to the proximal end of the catheter. At least 2 aspirations are performed. After removing the aspiration catheter, the Microseal adapter is reattached to the Guardwire and the occlusion balloon deflated allowing normal vessel flow. If the angiographic result is satisfactory, the device is removed.
• With filters, angiographic control is performed after stent deployment. If the result is satisfactory, the filter is withdrawn with the removal sheath. To facilitate protection device withdrawal and prevent it from getting caught on the stent, the guiding catheter is routinely advanced into the stent up its distal end. The aspirated blood or the filters are sent to the laboratory for analysis.

In some cases with filters, we can observe a slow or no-flow after stent deployment due to a clogged filter. In that case, there is a stagnant column of blood proximal to the filter. Simply recapturing the filter will result in embolization to the kidney. It is indispensable to perform an aspiration proximal to the filter prior to recapturing the device, either with an aspiration catheter or simply with the guiding catheter. Some aspiration catheters that can be used include the Export Catheter (Medtronic), Pronto (Vascular Solutions, Inc, Minneapolis, Minnesota), or the Diver (Invatec, Roncadello, Italy).

This technique of distal protection has some limitations we will discuss later.


To overcome some problems encountered with current filters, a new filter was recently proposed and is being investigated. This filter, the FiberNet® (Lumen Biomedical Inc, Plymouth, Minnesota), consists primarily of polyester fibers located coaxially around the distal tip of a guidewire assembly (Figure 2).

This filter is soft, conformable and when actuated, expands radially to fill the vessel providing excellent apposition to the vessel wall. Contained and captured emboli are recovered/removed both by aspiration through the retrieval catheter, and also by retention within the filter fibers when the filter is closed and retracted into the retrieval catheter. Aspiration is achieved through the retrieval catheter using the vacuum syringes to provide suction.

This filter enables capture emboli as small as 30/40 microns without compromising the flow through the filter, and the number of particles removed appeared much higher than with other filters.
The possibility of suction through the retrieval catheter during device removal is probably one of the major improvements with this device, allowing cleaning of the dilated area and of the inner part of the stent.
This technique allows aspiration of the debris, which can protrude through the struts of the stents after stent placement and dilatation and which could embolize after the procedure. The maximum diameter of this device is 7 mm, which is adequate in most of the cases. The landing zone required to place this filter is shorter than with other devices, about 1 cm.

We performed the first human study in renal arteries with this new protection device and the results will be published soon.

Medications and Patient Surveillance
In all of these stenting procedures, patients were given ticlopidine (500 mg/day) or clopidogrel (75 mg/day) and aspirin (100 mg/day) before the procedure.

During the procedure, an intravenous bolus of 5000 to 10,000 units of unfractioned heparin was routinely administrated at the beginning of the procedure to have an activated clotting time around 250–300 sec.
The post-procedural drug regimen included aspirin (100 mg/day) indefinitely and ticlopidine (250 mg/day) or clopidogrel (75 mg/day) for one month. Patients remained in the hospital for 48 hours to monitor serum creatinine levels and adjust blood pressure medication. Renal duplex scanning was scheduled the day after the procedure, at 6 and 12 months and then annually. Angiography was performed when a restenosis was suspected on the basis of positive clinical and duplex scan findings. Serum creatinine values are measured before and after the procedure (day 1) and at 1 and 6 months, with biannual measurements thereafter. For our last patients we evaluated the Glomerular Filtration Reserve (GFR), which seems a more reliable parameter to appreciate the renal function.

Results
Only single-center series on renal angioplasty and stenting under protection have been reported.

Personal series.
From January 1999 to July 2006, 124 RAS were treated in 105 patients with poorly controlled hypertension (72 males and 33 females), mean age was 64.5 ± 11.7 years (22–87), with percutaneous angioplasty and stenting under distal protection. All of these patients were diagnosed to have atherosclerotic RAS by renal duplex scanning and angiography.

The indication for endovascular treatment was a stenosis >70%.

Written informed consent was obtained from all patients.
One bilateral procedure was performed in 18 patients and we treated 2 renal arteries on the same side in one patient. Nine patients had a solitary or single functioning kidney (1 transplant renal artery). Twenty-four patients had moderate renal insufficiency (serum creatinine 1.5–1.9 mg / dl) and 15 had severe renal dysfunction (serum creatinine greater than or equal to 2 mg / dl).

The RAS was located at the ostium in 108 cases. Mean percentage stenosis was 85.9 ± 8.3% (70–99). Mean lesion length was 11.3 ± 3.0 mm (8–29).

The diameter of the artery was estimated at 6 mm in 87 arteries and 5 mm in 37 arteries. A total of 59 patients had diffuse severe atherosclerosis of the abdominal aorta. Of the patients, 32 had diabetes mellitus, 73 were current smokers, 66 had hyperlipidemia, and 73 had associated coronary disease. Cerebrovascular disease was found in 20 and lower extremity peripheral artery disease in 37.

A 6–8 Fr guiding catheter was placed at the ostium of the renal artery via a percutaneous femoral approach in all patients except one who presented with total occlusion of both iliac arteries necessitating a brachial approach.
Different protection devices were used:
• Occlusion balloon (Percusurge): 46 procedures
• Filters: 76 procedures
- EPI (Boston Scientific, Natick, Massachusetts): 59 (Figure 1)
- Angioguard (Cordis, Warren, New Jersey): 7
- Emboshield (Abbott Vascular, Redwood City, California): 6
- FiberNet (Lumen Biomedical, Plymouth, Minnesota): 4
- Accunet (Guidant, Indianapolis, Indiana): 2


Stenoses were easily crossed with the protection device in 123 cases due to their low profile and flexibility.
One predilatation was required to cross a subocclusive calcified stenosis. We had no difficulty in deflating the occlusion balloons or in removing the protection devices.

A direct stenting was performed in 96 cases, and 127 different stent models were used, including Boston Express (n = 27), Cordis Genesis (n = 29), Medtronic (n = 15), Guidant Herculink (n = 14), Cordis Corinthian (n = 11), Carbostent Sorin (n = 7), Cordis P 154 (n = 6), Abbott (n = 6), Nir (n = 5), Stentec (n = 4), Cordis M3 (n = 2), Biotronik (n = 1). Three patients required 2 stents in the same artery to treat a long lesion.
Technical success was obtained for all arteries (100%), with good stent deployment, no significant residual stenoses and complete covering of the lesion.

Mean renal artery occlusion time with the Percusurge technique was 6.46 ± 2.42 mm.
The mean time in situ for filters was 4.22 ± 1.18 mm, which is shorter than with the occlusion balloon.
Two patients developed an arterial spasm (one with Percusurge, one with EPI filter) at the site of protection device, which immediately responded to local vasodilatator therapy. No dissection of the renal artery due to a protection device was seen.

Particulate analysis.
• With the Percusurge technique and the aspiration catheter, we aspirated visible debris in all patients (as with the carotid procedures). The aspirated blood samples were analyzed and studied by microscopy and scanning electron microscopy. Different particles were isolated and identified. Their number varied from 13 to 208 per procedure (mean ± SD 98.1 ± 60.0), and diameter ranged from 38 to 6206 µm (mean ± SD 201.2 ± 76.2 µm). The particles were atheromatous plaques, cholesterol crystals, necrotic cores, fibrin, fresh thrombi, organized thrombi, platelets, and macrophage foam cells, thrombogenic lipoid masses. Only fresh thrombi should be drug sensitive but these emboli were detected in only 10% of cases. The other particles are likely to be manageable only using mechanical means.
• We removed, with filters, visible debris in 80% of the cases. Two filters were totally blocked by large particles, with flow being totally interrupted, and 2 other filters were almost totally blocked with low flow. The flow was restored after aspiration with the guiding catheter and removal of the filter. One patient was 26 years old.
• With the new filter FiberNet® visible debris were removed in all cases and 5 times more than with other filters. Minimum length 28 microns, average length 105 microns, maximum length 6.8 mm. Of the particles, 70% were less than 100 microns. Average area for FiberNet® debris captured: 117.1 mm2 (25.1–208). Atheromatous material was found in both aspirate and filter samples (65% in aspirate samples).


Follow up. The mean follow-up period was 18.3 ± 9 months (2–89 months). Four patients died from myocardial infarction, one at day 3 (patient who underwent coronary angioplasty shortly before renal stenting), and one at 6 months and one at 1 year of follow-up. One multivascular patient with severe renal insufficiency developed an acute renal failure after the procedure probably due to contrast nephropathy necessitating dialysis, and died one week later from multiorgan failure. Two patients were lost to follow up after one year, and 7 patients presented an in-stent restenosis successfully treated by a new balloon angioplasty. We have not seen device-related late vascular lesions in the 38 angiographic controls performed during the follow up.

Hypertension. During the follow up we observed no difference with other published series of patients treated without protection. Systolic and diastolic blood pressure decreased significantly. The number of medications also declined significantly (2.8 vs. 1.2).

Nineteen patients were cured (18%), 61 improved (59%), and 25 remained unchanged (22%).

Renal function. As mentioned, we observed one acute function deterioration. Table 2 shows the mean value for urea and creatinine preprocedurally (D-1) and after the procedure (D+1), at one month, 6 months, 1 year and 3 years. There is no statistically significant difference during the follow-up.


Table 3 shows the effects on the renal function (RF) at 6 months, 2 and 3 years in patients with normal renal function and in patients with moderate and severe renal insufficiency.

At 6 months, (91 patients), we observed only one deterioration of the RF in a patient with moderate renal insufficiency at baseline, 21 improvements in patients with renal insufficiency, 69 stabilizations.
At 2 years (75 patients), we observed only 2 deteriorations of the RF (3%):

• One in a patient with normal RF before the procedure, normal RF at 6 month follow-up and who developed a renal insufficiency at 2 years. This patient was treated for bilateral stenosis, one side without protection.
• One in a patient with moderate renal insufficiency. 96% of the patients remained stabilized (n = 54) or improved (n = 19).
At 3 years (37 patients), 93% of the patients remained stabilized (n = 26) or improved (n = 9). Twenty patients have RF deterioration.

Other published series
Few reports have been published in the literature. Holden et al reported a first series of 46 procedures in 37 patients with preprocedural renal impairment performed with the Angioguard filter.66 They found the same results: renal function stabilized or improved in 95% of cases. Only 5% of the patients demonstrated an unchanged decline. No patients experienced acute post procedural deterioration. These results are better than in most reports in the literature and also better than in a group of similar patients with ischemic nephropathy who underwent stent revascularization without distal protection at the same institution.55 The improved results are thought to be due to the prevention of cholesterol embolization during the procedure by distal filter baskets. A total of 65% of the filters contained embolic material, including fresh thrombus, chronic thrombus, atheromatous fragments and cholesterol clefts. More recently, Holden et al80 published a larger series of 106 RAS treated under protection with filters in 90 patients with ischemic nephropathy. He reported one acute deterioration of RF. At a mean follow up of 18.2 months, RF was improved in 36% of cases, stabilized in 55%. He reported only 8% of progressive decline of RF.

Eggebrecht et al published a case of renal angioplasty performed under distal protection with the Percusurge device for an in-stent restenosis.81 Histologic examination of blood retrieved from the distally occluded vessel showed foam cells and an amorphous lipoid substance as markers of atherosclerotic plaque debris. The authors concluded that the Guardwire could be included in routine angioplasty maneuvers.
Li and colleagues also published a case of successful renal angioplasty and stenting under distal protection with the Percusurge Guardwire.82

Guerkens and colleagues used the Angioguard filters in six patients and retrieved macroscopic debris in every case.83

Edwards et al84 more recently used balloon occlusion and an aspiration catheter (Percusurge device) to treat 32 RAS. Renal insufficiency was experienced by 92% of the patients. RF response at 4–6 week follow-up was improved in 50% of the patients and unchanged in 50%. In no patient or after any procedure was RF observed to be worsened, and 54% of the patients with RF deterioration were improved. For the authors, the results with this technique of RAS under protection represent a marked improvement in short-term RF response rates compared with previously published experiences and approximate the short-term results reported after open surgical revascularization, and they concluded that these data suggest that this technique may prevent RF harm during RAS as a result of atheroembolism and warrant further investigation.


1. Rihal CS, Textor SC, Breen JF, et al. Incidental renal artery stenosis among a prospective cohort of hypertensive patients undergoing coronary angiography. Mayo Clin Proc 2002;77:309–316.
2. Missouris CG, Buckenham T, Cappucio FP, et al. Renal artery stenosis: A common and important problem in patients with peripheral vascular disease. Am J Med 1994;96:10–14.
3. Gross CM, Krämer J, Waigland J, et al. Ostial renal artery stent placement for atherosclerotic renal artery stenosis in patients with coronary artery disease. Cath Cardiovasc Diagn 1998;45:1–8.
4. Jean WJ, al-Bitar I, Zwicke DL, et al. High incidence of renal artery stenosis in patients with coronary artery disease. Cath Cardiovasc Diagn 1994;32:8–10.
5. O’Neil EA, Hansen KJ, Canzanello VJ, et al. Prevalence of ischemic nephropathy in patients with renal insufficiency. Am Surg 1992;58:485–490.
6. Hansen KJ, Edwards MS, Craven TE, et al. Prevalence of renovascular disease in the elderly: A population based study. J Vasc Surg 2002;36:443–451.
7. Schreiber MJ, Pohl MA, Novick AC. The natural history of atherosclerotic and fibrous renal artery disease. Uro Clin North Am 1984;11:383–392.
8. Strandness DE Jr. Natural history of renal artery stenosis. Am J Kidney Dis 1994;24:630–635.
9. Zierler RE. Screening for renal artery stenosis: Is it justified? Mayo Clin Proc 2002;77:307–308.
10. Subramanyan R, Joy J, Balakrishnan KG. Natural history of aortoarteritis (Takayasu’s disease). Circulation 1989;80:429–437.
11. Sharma S, Jain AK. Natural history of non-specific aortoarteritis. In Panja M (ed). Non-specific Aorto Arteritis. Marksman Media Services Kolkata (India) 2001:p.165.
12. Goncharenko V, Gerlock AJ, Schaff MI, et al. Progression of renal artery fibromuscular dysplasia in 42 patients as seen on Angiography. Radiology 1981;139:45–50.
13. Knutson DW, Abt AB. Pathophysiology, pathology and clinical features of renovascular hypertension. In D Eugene Strandness, JR Arena Van Breda (eds): Vascular Diseases. New York: Churchill Livingstone Inc., 1994:p.673.
14. Berglund G, Anderson O, Wilhelmsent. Prevalence of primary and secondary hypertension studies in a random population sample. BMJ 1976;2:pp.554–556.
15. Pickering TG. Renovascular hypertension: Medical evaluation and non surgical treatment. In Laragh JM, Brenner BM (eds): Hypertension: Physiopathology, Diagnosis and Management. New York: Raven Press, 1990:p1539.
16. Messena LM, Zelenock GB, Yaok A, et al. Renal vascularization for recurrent pulmonary oedema in patients with poorly controlled hypertension and renal insufficiency: A distinct subgroup of patients with atherosclerotic renal artery occlusive disease. J Vasc Surg 1994;20:76–87.
17. Stane Y, Gyepes MT, Olson DL, et al. Renovascular hypertension in children and adolescents. Radiology 1978;129:123–128.
18. Sos TA, Pickering TG, Sniderman K, et al. Percutaneous transluminal angioplasty in renovascular hypertension due to atheroma or fibromuscular hyperplasia. N Engl J Med 1983;309:274–278.
19. Edwards MS, Craven TE, Borke GL, et al. Renovascular disease and the risk of adverse coronary events in the elderly — A prospective, population based study. Arch Intern Med 2005;165:207–213.
20. Fried LF, Shlipak MG, Crump C, et al. Renal insufficiency as a predictor of cardiovascular outcomes and mortality in elderly individuals. J Am Coll Cardiol 2003;41:1364–1372.
21. Johansson M, Herlitz H, Jensen G, et al. Increased cardiovascular mortality in hypertensive patients with renal artery stenosis. Relation to sympathic activation, renal function and treatment regimens. J Hypertens 1999;17:1743–1750.
22. Shlipak MG, Fried LF, Crump C, et al. Cardiovascular disease risk status in elderly persons with renal insufficiency. Kidney Int 2002; 62:997–1004.
23. Shlipak MG, Fried LF, Crump C, et al. Elevations of inflammatory and procoagulant biomarkers in elderly persons with renal insufficiency. Circulation 2003;107:87–92.
24. Novick AC. Ziegelbaum M, Vidt DG, et al. Trends in surgical revascularization for renal artery disease: Ten year’s experience. JAMA 1987;257:498–501.
25. Weibull H, Bergqvist D, Bergentz SE, et al. P.T.R.A. versus surgical reconstruction of atherosclerotic renal stenosis: A prospective randomized study. J Vasc Surg 1993;18:841–852.
26. Cambria RP. Surgery: Indications and variables that affect procedural outcome, as well as morbidity and mortality. J Invasive Cardiol 1998;10:55–58.
27. Van de Ven PJ, Kaatee R, Beutler JJ, et al. Arterial stenting and balloon angioplasty in ostial atherosclerotic renovascular disease: A randomized trial. Lancet 1999;353:282–286.
28. Henry M, Amor M, Henry I, et al. Stents in the treatment of renal artery stenosis: Long-term follow-up. J Endovasc Surg 1999;6:42–51.
29. Henry M, Amor M, Henry I, et al. Stent placement in the renal artery: Three-year experience with the Palmaz stent. J Vasc Interv Radiol 1996;7:343–350.
30. Blum U, Krumme B, Flugel P, et al. Treatment of ostial renal artery stenosis with vascular endoprotheses after unsuccessful balloon angioplasty. N Engl J Med 1997;336:459–465.
31. Van de Ven PJ, Beutler JJ, Kaatee R, et al. Transluminal vascular stent for ostial atherosclerotic renal artery stenosis. Lancet 1995;346:672–674.
32. Dorros G, Jaff M, Jain A, et al. Follow-up of primary Palmaz-Schatz stent placement for atherosclerotic renal artery stenosis. Am J Cardiol 1995;75:1051–1055.
33. Isles CG, Robertson S, Hill D. Management of renovascular disease: A review of renal artery stenting in ten studies. Q J Med 1999;92:159–167.
34. Boisclair C, Therasse E, Olivia VL, et al. Treatment of renal angioplasty failure by perculaneous renal artery stenting with Palmaz stents: Midterm technical and clinical results. AJR Am J Roentgenol 1997;168:245–251.
35. Hiramoto J, Hansen KJ, Pan WM et al. Atheroemboli during renal artery angioplasty: An ex vivo study. J Vasc Surg 2005;41:1026–1030.
36. Henry M, Klonaris C, Henry I. Renal stenting with the Percusurge Guardwire device: A pilot study. J Endovasc Ther 2001;8:227–237.
37. Henry M, Henry I, Klonaris C, et al. Renal angioplasty and stenting under protection. The way for the future? Catheter Cardiovasc Interv 2003;60:299–312.
38. Henry M, Henry I, Polydorou A, et al. Renal angioplasty and stenting: Long-term results and the potential role of protection devices. Expert Review Cardiovasc Ther 2005;3:321–334.
39. Nordmann AJ, Woo K, Parkes R, Logan AG. Balloon angioplasty or medical therapy for hypertensive patients with atherosclerotic renal artery stenosis? A meta-analysis of randomized controlled trials. Am J Med 2003;114:44–50.
40. Rundback JH, Sacks D, Kent KC, et al. Guidelines for the reporting of renal artery revascularisation in clinical trials. American Heart Association. Circulation 2002;106:1572–1585.
41. Iannone LA, Underwood P, Nath A, et al. Effect of primary balloon expandable renal artery stents on long-term patency, renal function and balloon pressure in hypertension and renal insufficient patients with renal artery stenosis. Cathet Cardiovasc Diagn 1996;37:243–250.
42. Taylor A, Sheppard D, MacLeod MJ, et al. Renal artery stent placement in renal artery stenosis: Technical and early clinical results. Clin Radiol 1997;52:451–457.
43. Harden PN, MacLeod MJ, Rodger RS, et al. Effect of renal artery stenting on progression of renovascular renal failure. Lancet 1997;349:1133–1136.
44. Paulsen D, Klow NE, Rogstad B, et al. Preservation of renal function by percutaneous transluminal angioplasty in ischemic renal disease. Nephrol Dial Transplant 1999;4:1454–1461.
45. Rodriguez-Lopez JA, Werber A, Ray L, et al. Renal artery stenosis treated with stent deployment: Indications, technique and outcome in 108 patients. J Vasc Surg 1999;29:617–624.
46. Rundback JH, Manoni T, Rozenblit GN, et al. Balloon angioplasty or stent placement in patients with azotemic renovascular disease: A retrospective comparison of clinical outcomes. Heart Dis 1999;1:121–125.
47. Guerrero M, Kunjummen B, Khaleel R, et al. Stabilization of renal function after renal artery stenting. Am J Cardiol 2002;90:63H.
48. Allaqaband S, Shaikh F, Zaidi S, et al. Long term effects of renal artery stenting or renal function in patients with atherosclerotic renal artery stenosis and baseline creatinine of > 2.0 mg/dl (abstract). Am J Cardiol 2003;92(suppl):54 L.
49. Haller ST, Kennedy D, Khuder S et al. Acute renal injury with renal artery stenting (abstract). Circulation 2004;110(suppl):III–614.
50. Zeller T, Frank U, Muller C, et al. Stent supported angioplasty of severe atherosclerotic renal artery stenosis preserves renal function and improves blood pressure control. J Endovasc Ther 2004;11:95–106.
51. Dorros G, Jaff M, Mathias L, et al. Multicenter Palmaz Stent renal artery stenosis revascularization registry report: Four-year follow-up of 1058 successful patients. Catheter Cardiovasc Interv 2002;55:182–188.
52. Fiala LA, Jackson MR, Gillespie DL, et al. Primary stenting of atherosclerotic renal artery ostial stenosis. Ann Vasc Surg 1998;12:128–133.
53. Plouin PF, Chatellier G, Darne B, et al. Blood pressure outcome of angioplasty in atherosclerotic renal artery stenosis: A randomized trial. Essai Multicentrique Medicaments vs Angioplastie (EMMA) Study group. Hypertension 1998;31:823–829.
54. White CJ, Ramee SR, Collins TJ, et al. Renal artery stent placement: Utility in lesions difficult to treat with balloon angioplasty. J Am Coll Cardiol 1997;30:1445–1450.
55. Dorros G, Jaff M, Mathiak L, et al. Four-year follow-up of Palmaz-Schatz stent revascularization as treatment for atherosclerotic renal artery stenosis. Circulation 1998;98:642–647.
56. Rocha Singh KJ, Miskkel GJ, Kathodi RE, et al. Clinical predictors of improved long-term blood pressure control after successful stenting of hypertensive patients with obstructive renal artery atherosclerosis. Cathet Cardiovasc Intervent 1999;47:167–172.
57. Von Knorring J, Edgren J, Lepantalo M. Long-term results of percutaneous transluminal angioplasty in renovascular hypertension. Acta Radiol 1996;37:36–40.
58. Watson P, Madjipetrou P, Cow S, et al. Effects of renal artery stenting on renal function and size in patients with atherosclerotic renovascular disease. Circulation 2000;102:1671–1677.
59. Rocha Singh K. Renaissance trial: A prospective multicenter trial to confirm the safety and efficacy of the express SD stent. Presented at the ISET Meeting in Miami Beach, Florida, January 22–26, 2006.
60. La Batide-Alanore A, Azizi M, Froissart M, et al. Split renal function outcome after renal angioplasty in patients with unilateral renal artery stenosis. J Am Soc Nephrol 2001;12:1235–1241.
61. Pattynama P, Becker GJ, Brown J, et al. Percutaneous angioplasty for atherosclerotic renal artery disease: Effect on renal function in azotemic patients. Cardiovasc Intervent Radio 1994;17:143–146.
62. Subramanian R, Silva JA, Ramee SR, et al. Beneficial effects of chronic renal insufficiency. Euro Heart J 2002;P577:97.
63. Topol EJ, Yadav JS. Recognition of the importance of embolization in atherosclerotic vascular diseases. Circulation 2000;101:570–580.
64. Baim DS, Wahr D, George B, et al. Randomized trial of a distal embolic protection device during percutaneous intervention of saphenous vein aorto-coronary bypass grafts. Circulation 2002;105:1285–1290.
65. Limbruno U, Micheli A, De Carlo M, et al. Mechanical protection of distal embolization during primary angioplasty: Safety, feasibility, and impact on myocardial reperfusion. Circulation 2003;108:171–176.
66. Holden A, Hill A. Renal angioplasty and stenting with distal protection of the main renal artery in ischemic nephropathy: Early experience. J Vasc Surg 2003;38:962–968.
67. Walker C, Kowalski J, Knan M, et al. Proximal protection before distal protection: Preventing large atheroemboly during renal intervention. Am J Cardiol 2002;90:28H.
68. Nolan BW, Schermerhom ML, Rowell E, et al. Outcomes of renal artery angioplasty and stenting using low-profile systems. J Vasc Surg 2005;41:46–52.
69. Commeau P, Barragan P, Huret B, et al. Direct stenting in the treatment of renal artery stenosis: Immediate and in-hospital results. Am J Cardiol 2001;TCT 74.
70. Morice MC, Marco J, Laborde JC, et al. The French registry on renal stenting. Am J Cardiol 2002;90:67H.
71. Rocha Singh K, Jaff MR, Rosenfield K. Evaluation of the safety and effectiveness of renal artery stenting after unsuccessful balloon angioplasty: The Aspire 2 study. J Am Coll Cardiol 2005;46:776–783.
72. Thadani RI, Camargo CA Jr, Xavier RJ, et al. Atheroembolic renal failure after invasive procedures. Natural history based on 52 histologically proven cases. Medicine (Baltimore). 1995;74:350–358.
73. Haqqie SS, Urizar RE, Singh J. Nephrotic-range proteinuria in renal atheroembolic disease: Report of four cases. Am J Kidney Dis 1996;28:493–501.
74. Krishnamurthi V, Novick AC, Myles JL. Atheroembolic renal disease: Effect on morbidity and survival after renovascularization for atherosclerotic renal artery stenting. Catheter Cardiovasc Interv 1999; 46:245–248.
75. Boero R, Borcam M, Ladarola GM, et al. Acute kidney failure caused by cholesterol embolism. Minerva Urol Nephrol 2000;52:119–122.
76. Goldstein JA, Kolluri R, Rocha Singh K. Technical considerations for renal artery stenting. Vascular Disease Management 2006;3:260–265.
77. Henry M, Henry I, Klonaris C, et al. Benefits of cerebral protection during carotid stenting with the percursurge guardwire system: Midterm results. J Endovasc Ther 2002;9:1–13.
78. Reimers B, Corvaja N, Moshiris A, et al. Cerebral protection with filter devices during carotid artery stenting. Circulation 2001;104:12–15.
79. Roubin GS. Carotid angioplasty and stenting under cerebral protection: The standard of care. Presented at the International Congress XV in Scottsdale, Arizona, February 11–14, 2002.
80. Holden A. Renal angioplasty and stenting with distal protection. Presented at the All That Jazz Meeting in New Orleans, Louisiana, April 26–29, 2005.
81. Eggebrecht H, Haude M, Baumgart D, et al. A new temporary occlusion and aspiration system prevention of distal embolization during percutaneous transluminal renal angioplasty. J Intervent Cardiol 2000;13:339–342.
82. Li SS, Wang CH, Lam CW. Renal angioplasty under protection of the percusurge Guardwire Plus System. J Invasive Cardiol 2003;15:148–150.
83. Guerkens U, Mueller R, Soblik S, et al. A new protection device to prevent distal embolization in peripheral interventions: First results in renal and carotid arteries. Presented at the ACC Meeting in Anaheim, CA, March 12–15, 2000.
84. Edwards MS, Craven BL, Stafford J, et al. Distal embolic protection during renal artery angioplasty and stenting. J Vasc Surg 2006;44:128–135.
85. Sos T, Trost D. The importance of renal stenting in ischemic nephropathy. Endovascular Today 2003;2:44–46.
86. Kulwant SH, Venkateswara KR. Atheroemboli Renal Disease. J Am Soc Nephrol 2001;12:1781–1787.

Vascular Disease Management - ISSN: 1553-8036 - Volume 4 - Issue 3 - May 2007 - Pages: 99 - 111

VASCULAR TOPICS

Peripheral Angioplasty
Thoracic Stent Grafts
Renal Stenting
Vena Cava Filter
Hemodialysis Management
Computed Tomography
PFO Closure
SFA Stenting
Carotid Stenting
Vessel Closure
Angiography
Carotid Endarterectomy
Ultrasound

Critical Limb Ischemia
Superficial Femoral Artery
Embolization
Device Based Thrombectomy
Pharmacological Management
MRA
Mesenteric Artery Stenting
AAA Stent Grafts
Iliac Stenting
Thrombolysis Procedures Using Drug Therapy

SUPPLEMENTS

Superior Mesenteric Artery Revascularization and Retrograde Visualization
This clinical case update was supported through an unrestricted educational grant from Terumo Medical Corporation.

HMP Increased Cutaneous Sensibility in Patients with Diabetic Neuropathy Utilizing a Pharmacological Approach — Clinical Case Evidence

This clinical case update was supported through an unrestricted educational grant from Pamlab, LLC.

A New Biological Approachto Below-Knee Revascularization
A Review of the GORE PROPATEN Vascular Graft:
The Combination That Lasts

This special supplement was made possible through a grant from W. L. Gore

Combining Bilayered Living Cell Therapy with Minimally Invasive Vein Surgery:
Current Treatment Strategies for Venous Ulcers

This activity is supported by an educational grant from Organogenesis.

Pharmacotherapy in Peripheral Vascular Disease

Platelet Inhibition in Critical Limb Ischemia and Peripheral Vascular Interventions
DAVID E. ALLIE, MD

An Overview of Pharmacotherapy during Percutaneous Peripheral Interventions of Thrombotic Lesions
NICOLAS W. SHAMMAS, MD, MS, FACC


The Important Properties of Contrast Media: Focus on Viscosity

This special supplement was made possible through a grant from Guerbet LLC

RECENTLY ADDED

Anticoagulation Techniques for Peripheral Vascular Interventions

Archived Webcast





© 2008 HMP Communications | All Rights Reserved
83 General Warren Blvd., Suite 100, Malvern, PA 19355
Phone: 800.237.7285 610.560.0500 Fax: 610.560.0502
Privacy Policy